Class III Models

950mm

1360mm

Inflow air volume

168 m3/Hr

240 m3/Hr

Exhaust model pressure drop across the cabinet

350Pa clean filters
450Pa dirty filters

350Pa clean filters
450Pa dirty filters

Inlet HEPA filter type

99.997% @ 0.3 micron

99.997% @ 0.3 micron

Main HEPA filter type

99.997% @ 0.3 micron

99.997% @ 0.3 micron

Electric inputs
1.For fans,lights and control/alarm system
2.Internal electrical socket.

Recirculating model 240v 50Hz
Current usage is dependant on fan size.

Recirculating model 240v 50Hz
Current usage is dependant on fan size.

Electric outputs
1.For external exhaust fan

2.Volt free contacts

 
Power usage is duct length related.
0-240v speed controlled.
Optional extra for alarm system monitoring and external fan control.

 
Power usage is duct length related.
0-240v speed controlled.
Optional extra for alarm system monitoring and external fan control.

Lighting levels

Average 1000 Lux

Average 1000 Lux

Paint

Epoxy Polyester powder coated with a 40 micron minimum paint thickness.

Epoxy Polyester powder coated with a 40 micron minimum paint thickness.

Noise Levels

Usually 58dB or less

Usually 58dB or less

Weights

130kg

150kg


Dimensions  

A

All models 1110mm

B

All models Work area height 730mm

X

3ft Cab. 950mm
4ft Cab. 1360mm

Y

3ft Cab. 1070mm
4ft Cab. 1480mm