Class II Models
|
990mm
|
1290mm
|
1890mm
|
|
Inflow air velocity
|
0.55 m/s
|
0.55 m/s
|
0.55 m/s
|
|
Downflow air velocity
|
0.35 m/s
|
0.35 m/s
|
0.35 m/s
|
|
Air extract volume
|
356 m3/Hr
|
475 m3/Hr
|
713 m3/Hr
|
|
HEPA filter type
|
99.997% @ 0.3 micron
|
99.997% @ 0.3 micron
|
99.997% @ 0.3 micron
|
|
Electric inputs
1.For fans,lights and control/alarm system
2.Internal electrical socket.
|
Recirculating model 240v
4.5A 50Hz up to 5A
|
Recirculating model 240v
5A 50Hz up to 5A
|
Recirculating model 240v
8A 50Hz up to 5A
|
|
Electric outputs
1.For external exhaust fan
2.Volt free contacts
|
Power usage is duct length related.
0-240v speed controlled. Optional extra for alarm system
monitoring and external fan control.
|
Power usage is duct length related.
0-240v speed controlled. Optional extra for alarm system
monitoring and external fan control.
|
Power usage is duct length related.
0-240v speed controlled. Optional extra for alarm system
monitoring and external fan control.
|
|
Lighting levels
|
Average 1000 Lux
|
Average 1000 Lux
|
Average 1000 Lux
|
|
Paint
|
Epoxy Polyester powder coated
with a 40 micron minimum paint thickness.
|
Epoxy Polyester powder coated
with a 40 micron minimum paint thickness.
|
Epoxy Polyester powder coated
with a 40 micron minimum paint thickness.
|
|
Noise Levels
|
Usually 58dB or less for
recirculating models and 56dB or less for exhaust models.
|
Usually 58dB or less for
recirculating models and 56dB or less for exhaust models.
|
Usually 58dB or less for
recirculating models and 56dB or less for exhaust models.
|
|
Weights
|
125kg
|
255kg
|
315kg
|