Class II Models

990mm

1290mm

1890mm

Inflow air velocity

0.55 m/s

0.55 m/s

0.55 m/s

Downflow air velocity

0.35 m/s

0.35 m/s

0.35 m/s

Air extract volume

356 m3/Hr

475 m3/Hr

713 m3/Hr

HEPA filter type

99.997% @ 0.3 micron

99.997% @ 0.3 micron

99.997% @ 0.3 micron

Electric inputs
1.For fans,lights and control/alarm system
2.Internal electrical socket.

Recirculating model 240v 4.5A 50Hz up to 5A

Recirculating model 240v 5A 50Hz up to 5A

Recirculating model 240v 8A 50Hz up to 5A

Electric outputs
1.For external exhaust fan

2.Volt free contacts

 
Power usage is duct length related.
0-240v speed controlled. Optional extra for alarm system monitoring and external fan control.

 
Power usage is duct length related.
0-240v speed controlled. Optional extra for alarm system monitoring and external fan control.

 
Power usage is duct length related.
0-240v speed controlled. Optional extra for alarm system monitoring and external fan control.

Lighting levels

Average 1000 Lux

Average 1000 Lux

Average 1000 Lux

Paint

Epoxy Polyester powder coated with a 40 micron minimum paint thickness.

Epoxy Polyester powder coated with a 40 micron minimum paint thickness.

Epoxy Polyester powder coated with a 40 micron minimum paint thickness.

Noise Levels

Usually 58dB or less for recirculating models and 56dB or less for exhaust models.

Usually 58dB or less for recirculating models and 56dB or less for exhaust models.

Usually 58dB or less for recirculating models and 56dB or less for exhaust models.

Weights

125kg

255kg

315kg


Dimensions  

A

1440mm for Recirc.
1324mm for Exhaust

B

700mm for all

X

For 3ft Cab. 990mm
For 4ft Cab. 1290mm
For 6ft Cab. 1890mm

Y

For 3ft Cab. 900mm
For 4ft Cab. 1200mm
For 6ft Cab. 1800mm