You can EMail us from here will test your clean room to your own specifications. We recommend that each unit is checked once every 6 months for:-

1.

Operation of all services and system controls.

2.

Filter and seal integrity tests.

3.

Inspection of room for damage.

4.

Check inflow/outflows to calculate air change rates.

5.

Check alarm and airflow gauge for correct operation and calibration (if fitted).

6.

Pre filter changed if necessary.

7. Particle count all areas/rooms.

Air inflow measurement - Using - Airflow Developments AV2 Anemometer.
Horizontal/Vertical Laminar Flow unit - Measure airflow, a minimum of eight positions should be used and these should be recorded on the service report.

As per requirements BS5295:1989

Filter and seal integrity tests - Using - Taylor Smoke generator and ATI forward light scattering photometer
As per requirements of Appendix B of BS5726: Part 1:1992. Section 8.5.

Give full test certificate.
All test equipment where necessary is always calibrated and carries a full calibration test certificate.
Any of these certificates are available on special request.